7
Error Information
Detector Error Messages
Cutoff filter doesn't decrease the light intensity at 250 nm
Error ID: 7813
The automatic filter check after lamp ignition has failed. When the lamp is
switched on, the detector moves the cutoff filter into the light path. If the
filter is functioning correctly, a decrease in lamp intensity is seen. If the
expected intensity decrease is not detected, the error message is
generated.
Probable cause
1
2
3
4
ADC Hardware Error
Error ID: 7830, 7831
A/D- Converter hardware is defective.
Probable cause
1
100
Motor is not connected.
Defective motor.
Defective/missing grating or filter.
Cable/connector defective.
A/D-Converter hardware is defective.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Agilent 1260 Infinity VWD User Manual