Troubleshooting And Test Functions; Overview Of The Detector's Indicators And Test Functions - Agilent Technologies 1100 Series Reference Manual

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

3

Troubleshooting and Test Functions

Overview of the Detector's Indicators and Test Functions

Status Indicators
The detector is provided with two status indicators which indicate the
operational state (prerun, run, and error states) of the detector. The status
indicators provide a quick visual check of the operation of the detector (see
page
Error Messages
In the event of an electronic, mechanical or hydraulic failure, the detector
generates an error message in the user interface. For each message, a short
description of the failure, a list of probable causes of the problem, and a list of
suggested actions to fix the problem are provided (see
page 53).
Wavelength Recalibration
Wavelength recalibration is recommended after repair of internal
components, and on a regular basis to ensure correct operation of the
detector
wavelength calibration (see
Test Functions
A series of test functions are available for troubleshooting and operational
verification after exchanging internal components (see
page 85).
50
51).
detector
. The
uses the deuterium alpha and beta emission lines for
"Wavelength Recalibration"
1100 Series DAD and MWD Reference Manual
"Error Messages"
on
on page 84).
"Test Functions"
on