7
Error Information
Detector Error Messages

Grating or Filter Motor Errors

Error ID: Grating: 7800, 7801, 7802, 7803, 7804, 7805, 7806, 7808, 7809; Filter:
7810, 7811, 7812, 7813, 7814, 7815, 7816
The motor test has failed.
Test 0 Failed:
Test 1 Failed:
During the motor tests, the detector moves the motor to the end position while
monitoring the end-position sensor. If the end position is not found, the error
message is generated.
Probable cause
1
2
3
4

Wavelength test failed

Error ID: 7890
The automatic wavelength check after lamp ignition has failed. When the lamp is
switched on, the detector waits 1 min to warm-up the lamp. Then a check of the
deuterium emission line (656 nm) via the reference diode is performed. If the
emission line is more than 3 nm away from 656 nm, the error message is
generated.
Probable cause
1
Agilent InfinityLab LC Series VWD User Manual
Motor is not connected.
Defective motor.
Defective/missing grating or filter.
Cable/connector defective.
Calibration incorrect.
Filter motor.
Grating motor.
Suggested actions
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.
Suggested actions
Recalibrate the detector.