POLYDOROS SX
With XCU-HD unit 38 27 123
•
Remove the lithium battery from the CPU-HD unit (1a/Fig. 1).
With SX console unit 58 95 292
•
Remove the lithium battery from the slot CPU (1b/Fig. 1).
D320 CPU-HD unit
•
Remove the S-RAM J56 with the lithium battery from PC board D320 (4/Fig. 1).
High-voltage generator H1
•
Remove the high-voltage generator H1 from the generator cabinet.
•
Unscrew the cap screws from the high-voltage generator (2/Fig. 1).
•
Lift the cover (3/Fig. 1) with the attachments off the trough of the high-voltage generator
and pour the oil (approx. 25 l) into a suitable container.
ASPIA imaging system
Fig. 2
•
Remove the cover of the PCs on the left side of the ASPIA container (Fig. 2).
•
Dispose of the coin cell on the motherboard (1/Fig. 2); lithium battery.
UROSKOP ACCESS
System
Disassembly and Disposal
SPL5-330.861.01
Page 4 of 8
Rev. 02
04.05
CS PS 24