Technical specification of ADAM-4015
Channel Number
Support Protocol
Input Type
Input Connections
Wire Burnout Detection
Input Type and
Temperature Range
Isolation Voltage
Sampling Rate
Input Impedance
Resolution
Accuracy
CMR@50/60Hz
NMR@50/60Hz
Span Drift
Zero Drift
Watchdog Timer
Power Input
Power Consumption
ADAM ASCII and MODBUS/RTU
Pt100, Pt1000, BALCO500, Ni
2 or 3 wires
Yes
Pt100: -50 to 150° C
0 to 100° C
0 to 200° C
0 to 400° C
-200 to 200° C
Pt1000: -40 to 160° C
Balco500: -30 to 120° C
Ni 50 RTD: -80 to 100° C
Ni 508 RTD: 0 to 100° C
3000 V
DC
10 sample/second (total)
10 MΩ
16-bit
±0.1% or better
120 dB
100 dB
± 25 ppm/℃
± 3 μV/℃
System (1.6 second) and Communication
+10~+30 V
1.2 W @ 24V
Table 3-1 Technical specification of ADAM-4015