Advantech Data Acquisition Modules ADAM 4000 User Manual page 119

Advantech data acquisition modules user's manual
Table of Contents

Advertisement

Command Set
ADAM-4017+
Input range and external calibrating input source for each input type
Range
Input Range
bcode(Hex)
07
+/-4~20 mA
08
+/-10 V
09
+/-5 V
0a
+/-1 V
0b
+/-500 mV
0c
+/-150 mV
0d
+/-20 mA
4-16 ADAM 4000 Series User's Manual
Max. Signal
Min Signal
+20.000
-20.000
+10.000
-10.000
+5.0000
-5.0000
+1.0000
-1.0000
+500.00
-500.00
+150.00
-150.00
+20.000
-20.000
Displayed
Resolution
1uA
1mV
100.00uV
100.00uV
10uV
10uV
1uA

Advertisement

Table of Contents
loading