TRACe Subsystem
13

Error Messages

– If a P-Series sensor is not connected, error –241, "Hardware missing" occurs.
– If a P-Series sensor is connected and Free Run trigger acquisition is selected,
error –221, "Settings conflict" occurs.
TIf you attempt to measure a pulse out of the range of the capture, for example,
NOTE
measure the 5th pulse and there are only 4 pulses displayed, the power meter
returns #0##9.91E37 as the result.