Keysight Technologies P Series Programming Manual page 468

Modular power meter
Hide thumbs Also See for P Series:
Table of Contents

Advertisement

12
SYSTem Subsystem
–330
Self-test Failed;Measurement Channel B Fault
–330
Self-test Failed;Calibrator Fault
Refer to "Calibrator" on page 104 if you require a description of the calibrator test.
–330
Self-test Failed;ROM Check Failed
–330
Self-test Failed;RAM Check Failed
–330
Self-test Failed;Display Assy. Fault
Refer to "Display" on page 104 if you require a description of the Display test.
–350
Queue overflow
The error queue is full and another error has occurred which could not be recorded.
–361
Parity error in program
The serial port receiver has detected a parity error and consequently, data integrity cannot be guaranteed.
–362
Framing error in program
The serial port receiver has detected a framing error and consequently, data integrity cannot be
guaranteed.
–363
Input buffer overrun
The serial port receiver has been overrun and consequently, data has been lost.
–410
Query INTERRUPTED
A command was received which sends data to the output buffer, but the output buffer contained data from
a previous command (the previous data is not overwritten). The output buffer is cleared when power has
been off, or after *RST (reset) command has been executed.
–420
Query UNTERMINATED
The power meter was addressed to talk (that is, to send data over the interface) but a command has not
been received which sends data to the output buffer. For example you may have executed a CONFigure
command (which does not generate data) and then attempted to read data from the remote interface.
–430
Query DEADLOCKED
A command was received which generates too much data to fit in the output buffer and the input buffer is
also full. Command execution continues but data is lost.
–440
Query UNTERMINATED after indefinite response
The *IDN? command must be the last query command within a command string.
468
Keysight N8262A Programming Guide

Advertisement

Table of Contents
loading

This manual is also suitable for:

N8262a

Table of Contents