MEMory Subsystem
Syntax
MEM
Parameters
Item
numeric_value
Example
MEM:TABL:SEL "Sensor_1"
MEM:TABL:GAIN 97,99.5,97.4
Query
MEMory:TABLe:GAIN[:MAGNitude]?
The query returns a list of calibration factor/offset points for the currently selected
table.
310
:GAIN
:TABL
Description/Defaul t
A numeric value for the calibration/ offset factors. The units
are PCT.
Space
:MAGN
Range of Values
1.0 to 150.0
This command enters a reference calibration
factor of 97 % and calibration factors of 99.5
% and 97.4 % into the sensor calibration
table.