[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO <boolean>
This command is only valid when used with an E-Series Power Sensor. Its purpose
is to enable and disable autoranging. When autoranging is ON, the power meter
selects the best measuring range for the measurement. When autoranging is set
to OFF, the power meter remains in the currently set range.
The [SENSe[1]]|SENSe2:POWer:AC:RANGe command disables autoranging.
If INITiate:CONTinuous is set to ON and TRIGger:SOURce is set to IMMediate, the
range tracks the input power if [SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO is ON.
If the power meter is not making measurements then autoranging only occurs
when the power meter is triggered.
Syntax
SENS
1
SENS2
Example
POW:AC:RANG:AUTO 0
Reset Condition
On reset, autoranging is enabled.
Keysight N8262A Programming Guide
:
POW
:AC
:RANG
:AUTO
This command disables autoranging.
SENSe Subsystem
Space
0|OFF
1|ON
?
10