PSTatistic[1]|2:CCDF:TABle?
This command is used to return the data in CCDF table, average input power,
probability at the average input power, peak to average power ratio and sample
count.
This command will return 10 scalar results in the following order:
NOTE
1 Average input power (in dBm)
2 Probability at the average input power (in %)
3 Power level (power to average power ratio) that has 10 % of the power (in dB)
4 Power level (power to average power ratio) that has 1 % of the power (in dB)
5 Power level (power to average power ratio) that has 0.1 % of the power
6 Power level (power to average power ratio) that has 0.01 % of the power
7 Power level (power to average power ratio) that has 0.001 % of the power
8 Power level (power to average power ratio) that has 0.0001 % of the power
9 Peak to average power ratio (in dB)
10 Sample count
This command is only applicable when P-Series sensors are present and the
NOTE
following conditions are met:
– Free run acquisition mode is selected
– NORMal or DOUBle measurement speed setting is chosen
Keysight N8262A Programming Guide
(in dB)
(in dB)
(in dB)
(in dB)
PSTatistic Subsystem
9
337