For most applications, automatic filter length selection
NOTE
([SENSe[1]]|SENSe2:AVERage:COUNt:AUTO ON) is the best mode of operation.
However, manual filter length selection ([SENSe[1]]|SENSe2:AVERage:COUNt
<numeric_value>) is useful in applications requiring either high resolution or fast
settling times, where signal variations rather than measurement noise need
filtering, or when approximate results are needed quickly.
Syntax
SENS
SENS2
Example
AVER:COUN:AUTO OFF
Reset Condition
On reset, automatic averaging is enabled.
Query
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO?
The query enters a 1 or 0 into the output buffer indicating whether automatic filter
length is enabled or disabled.
– 1 is returned when automatic filter length is enabled
– 0 is returned when automatic filter length is disabled
Keysight N8262A Programming Guide
:
1
AVER
:COUN
:AUTO
This command disables automatic filter
length selection for Channel A.
SENSe Subsystem
Space
0|OFF
1|ON
ONCE
?
10
351