CALibration[1]|2:ZERO:NORMal:AUTO <boolean>
This command can only be used to zero a P-Series sensor.
The P-Series sensor only has a NORMAL path.
NOTE
The command causes the power meter to perform its zeroing routine, on the
specified channel, when ONCE is selected. This adjusts the power meter for a zero
power reading with no power supplied to the power sensor.
The 0|OFF parameter is only required for the query response and is ignored in the
command.
Syntax
CAL
Example
CAL2:ZERO:NORM:AUTO ONCE
Reset Condition
On reset, automatic zeroing is disabled.
Keysight N8262A Programming Guide
:ZERO
1
2
:NORM
:AUTO
This command causes the power meter to
perform a zeroing routine on Channel B.
CALibration Subsystem
Space
0|OFF
ONCE
?
4
239