Keysight Technologies P Series Programming Manual page 386

Modular power meter
Hide thumbs Also See for P Series:
Table of Contents

Advertisement

10
SENSe Subsystem
[SENSe[1]]|SENSe2:FREQuency[:CW|:FIXed] <numeric_value>
This command is used to enter a frequency. If the frequency does not correspond
directly to a frequency in the sensor calibration table, the power meter calculates
the calibration factor using linear interpolation. For 8480 Series Power Sensor the
power meter uses linear interpolation to calculate the calibration factor for the
frequency entered if [SENSe[1]]|SENSe2:CORRection:CSET:STATe is ON. For
P-Series and E-Series Power Sensor, the appropriate corrections are applied for
the frequency selected, dependant on the calibration data stored in the sensor's
EEPROM.
Syntax
SENS
SENS2
386
:
1
FREQ
Space
:CW
:FIX
?
Keysight N8262A Programming Guide
numeric_value
DEF
MIN
MAX
Space
MIN
MAX

Advertisement