Keysight Technologies P Series Programming Manual page 387

Modular power meter
Hide thumbs Also See for P Series:
Table of Contents

Advertisement

Parameters
Item
numeric_value
[a] The following measurement units can be used:
Hz
3
kHz (10
)
6
MHz (10
)
9
GHz (10
)
Example
FREQ 500kHz
Reset Condition
On reset, the frequency is set to 50 MHz (DEF).
Query
[SENSe[1]]|SENSe2:FREQuency[:CW|:FIXed]? [MIN|MAX]
The query returns the current frequency setting or the values associated with MIN
and MAX. The units in which the results are returned are Hz.
Query Example
SENS2:FREQ?
Keysight N8262A Programming Guide
Description/Defaul t
A numeric value for the frequency:
– DEF: the default value is 50 MHz
– MIN: 1 kHz
– MAX: 1000.0 GHz
The default units are Hz.
Range of Values
1 kHz to 1000.0 Ghz
DEF
MIN
MAX
This command enters a Channel A frequency
of 500 kHz.
This command queries the Channel B
frequency setting.
SENSe Subsystem
[a]
10
387