Power Meter Remote Operation
1
SENSe1:FREQuency 500KHZ
INITiate1:IMMediate
FETCh1?
READ? Example
ABORt1
CONFigure1:POWer:AC DEF,2,(@1)
SENS1:CORR:CSET2:SEL "Offset1"
SENS1:CORR:CSET2:STAT ON
SENSe1:FREQuency 500KHZ
READ1?
If the measurement frequency does not correspond directly to a frequency in the
NOTE
frequency dependent offset table, the power meter calculates the offset using
linear interpolation.
If you enter a frequency outside the frequency range defined in the frequency
dependent offset table, then the power meter uses the highest or lowest
frequency point in the table to set the offset.
To find out the value of the offset being used by the power meter to make a
measurement, use the query command:
SENSe:CORRection:GAIN4|FDOFfset[:INPut][MAGNITUDE]?
The response may be an interpolated value.