Measure[1]|2|3|4[:Scalar][:Power:ac]:Difference: Relative? [<Expected_Value>[,<Resolution>[,<Source List>]]]; Syntax; Parameters - Keysight Technologies P Series Programming Manual

Modular power meter
Hide thumbs Also See for P Series:
Table of Contents

Advertisement

2
MEASurement Commands
MEASure[1]|2|3|4[:SCALar][:POWer:AC]:DIFFerence:
RELative? [<expected_value>[,<resolution>[,<source list>]]]
This command sets the specified window's measurement function to difference
mode with relative mode on, aborts, configures then initiates both Channel A and
B, calculates the difference measurement result and places the result on the bus.
The result is a ratio based measurement and is expressed in the units defined by
UNIT[1]|2|3|4:POWer:RATio. The relative value used is that set by the
CALCulate:RELative:MAGNitude:AUTO command.

Syntax

MEAS
1
2
3
4

Parameters

Refer to
parameters in this command.
Item
expected_value
(for the expected power
level)
178
:SCAL
:POW
Space
expected_value
DEF
"Optional Parameters"
Description/Defaul t
The power meter ignores the numeric value entered in this
parameter. Any value entered is treated like DEF.
:AC
:DIFF
,
resolution
DEF
on page 126 for additional details on the
Keysight N8262A Programming Guide
:REL
?
,
source list
Range of Values
Sensor dependent
[a]
DEF