Calculate[1]|2|3|4:Limit:fail - Keysight Technologies P Series Programming Manual

Modular power meter
Hide thumbs Also See for P Series:
Table of Contents

Advertisement

CALCulate[1]|2|3|4:LIMit:FAIL?

This query enters a 1 or 0 into the output buffer indicating whether there have
been any limit failures for the specified window. A limit failure is defined as
CALC[1]|2|3|4:LIMit:FCO? being non-zero. The FCO (fail counter) can be zeroed
using the CALC[1]|2|3|4:LIMit:CLEar command.
– 1 is returned when one or more limit failures have occurred
– 0 is returned when no limit failures have occurred
Syntax
Example
CALC1:LIM:FAIL?
Reset Condition
On reset, the buffer is set to zero for both upper and lower window measurements.
Keysight N8262A Programming Guide
CALC
1
2
3
4
:LIM
:FAIL
This command queries if there have been any
limit failures on the upper window/upper
measurement.
CALCulate Subsystem
?
3
201