SYSTem Subsystem
–226
Lists not same length
This occurs when SENSe:CORRection:CSET[1]|CSET2:STATe is set to ON and the frequency and
calibration/offset lists do not correspond in length.
–230
Data corrupt or stale;Please calibrate Channel B
When CAL[1|2]:RCAL is set to ON and the sensor currently connected to Channel B has not been
calibrated, then any command which would normally return a measurement result (for example FETC?,
READ?, or MEAS?) will generate this error message.
–231
Data questionable;CAL ERROR
Power meter calibration failed. The most likely cause is attempting to calibrate without applying a 1 mW
power to the power sensor.
–231
Data questionable;CAL ERROR ChA
Power meter calibration failed on Channel A. The most likely cause is attempting to calibrate without
applying a 1 mW power to the power sensor.
–231
Data questionable;CAL ERROR ChB
Power meter calibration failed on Channel B. The most likely cause is attempting to calibrate without
applying a 1 mW power to the power sensor.
–231
Data questionable;Input Overload
The power input to Channel A exceeds the power sensor's maximum range.
–231
Data questionable;Input Overload ChA
The power input to Channel A exceeds the power sensor's maximum range.
–231
Data questionable;Input Overload ChB
The power input to Channel B exceeds the power sensor's maximum range.
–231
Data questionable;Lower window log error
This indicates that a difference measurement in the lower window has given a negative result when the
units of measurement were logarithmic.
–231
Data questionable;Upper window log error
This indicates that a difference measurement in the upper window has given a negative result when the
units of measurement were logarithmic.
–231
Data questionable;ZERO ERROR
Power meter zeroing failed. The most likely cause is attempting to zero when some power signal is being
applied to the power sensor.
–231
Data questionable;ZERO ERROR ChA
Power meter zeroing failed on Channel A. The most likely cause is attempting to zero when some power
signal is being applied to the power sensor.
466
Keysight N8262A Programming Guide