–168
Block data not allowed
A legal block data element was encountered but not allowed by the power meter at this point.
For example SYST:LANG #15FETC?.
–178
Expression data not allowed
A legal expression data was encountered but not allowed by the power meter at this point.
For example SYST:LANG (5+2).
–211
Trigger ignored
Indicates that <GET> or *TRG, or TRIG:IMM was received and recognized by the device but was ignored
because the power meter was not in the wait for trigger state.
–213
Init ignored
Indicates that a request for a measurement initiation was ignored as the power meter was already initiated.
For example, INIT:CONT ON
INIT.
–214
Trigger deadlock
TRIG:SOUR was set to HOLD or BUS and a READ? or MEASure?
was attempted, expecting TRIG:SOUR to be set to IMMediate.
–220
Parameter error;Frequency list must be in ascending order.
Indicates that the frequencies entered using the
MEMory:TABLe:FREQuency command are not in ascending order.
–221
Settings conflict
This message occurs under a variety of conflicting conditions. The following list gives a few examples of
where this error may occur:
If the READ? parameters do not match the current settings.
If you are in fast mode and attempting to switch on for example, averaging, duty cycle or limits.
Trying to clear a sensor calibration table when none is selected.
–222
Data out of range
A numeric parameter value is outside the valid range for the command.
For example, SENS:FREQ 2KHZ.
–224
Illegal parameter value
A discrete parameter was received which was not a valid choice for the command. You may have used an
invalid parameter choice.
For example, TRIG:SOUR EXT.
Keysight N8262A Programming Guide
SYSTem Subsystem
12
465