Table 1-4 Model Of Sensor And Measurement Rates; Trigger Mode - Keysight Technologies P Series Programming Manual

Modular power meter
Hide thumbs Also See for P Series:
Table of Contents

Advertisement

Sensor
Different measurement rates are achievable depending on the sensor type being
used, as shown in
Table 1-4
Model of sensor and measurement rates
Sensor
8480 Series
E-Series E4410 and E9300
P-Series

Trigger Mode

The power meter has a very flexible triggering system. For simplicity, it can be
described as having three modes:
– Free Run: When a channel is in Free Run, it continuously takes measurements
– Triggered Free Run: When a channel is in Triggered Free Run Continuous
– Single Shot: When a channel is in Single Shot, it takes a new measurement
Keysight N8262A Programming Guide
Table
Measurement Rate
NORMal
20 reading/s
50 ms
20 reading/s
50 ms
20 reading/s
on this channel. A channel is in free run when INITiate:CONTinuous is set to
ON and TRIGger:SOURce is set to IMMediate.
Trigger, it takes a new measurement each time a trigger event is detected. A
channel is in Triggered Free Run Continuous Trigger when
INITiate:CONTinuous is set to ON and TRIGger:SOURce is not set to
IMMediate.
when a trigger event is detected and then returns to the idle state. A channel is
in Single Shot when INITiate:CONTinuous is set to OFF. Note that a
measurement can take several INT/EXT triggers depending on the filter
settings. Refer to
"TRIGger[1]|2:DELay:AUTO <boolean>"
further information.
Power Meter Remote Operation
1-4:
DOUBle
40 reading/s
25 ms
40 reading/s
25 ms
40 reading/s
FAST
NA
Up to 400
Up to 1500
on page 580 for