Keysight Technologies P Series Programming Manual page 73

Modular power meter
Hide thumbs Also See for P Series:
Table of Contents

Advertisement

Power Meter Remote Operation
1
Some fine tuning of the above program segment can be carried out for example,
by setting the trigger delay off.
1 ABOR1
2 CONF1 DEF,3
3 TRIG1:DEL:AUTO OFF
4 INIT1:IMM
5 FETC1? DEF,3
Example 5 - Making a Difference Measurement
The following program segment can be carried out on the N8262A. It queries the
lower window to make a difference measurement of Channel A -Channel B. The
expected power level and resolution parameters are defaulted, leaving them at
their current settings. Some fine tuning of the measurement is carried out by
setting the averaging, and the trigger delay to off.
Using READ?
ABOR1
ABOR2
CONF2:POW:AC:DIFF DEF,DEF,(@1),(@2)
SENS1:AVER:COUN 1024
SENS2:AVER:COUN 1024
TRIG1:DEL:AUTO OFF
TRIG2:DEL:AUTO OFF
READ2:POW:AC:DIFF?
READ2:POW:AC:DIFF? DEF,DEF,(@2),(@1) (A second READ? query is sent to make
a Channel B -Channel A measurement using fresh measurement data).
Keysight N8262A Programming Guide
73