Service:bist:tbase:state <Boolean - Keysight Technologies P Series Programming Manual

Modular power meter
Hide thumbs Also See for P Series:
Table of Contents

Advertisement

16
SERVice Subsystem
SERVice:BIST:TBASe:STATe <boolean>
This command sends a 10 MHz time base signal to the front panel trig out for
testing purposes.
This command overrides the OUTPut:TRIGger[:STATe] command.
NOTE
For example, if OUTPut:TRIGger[:STATe] is ON and the command
SERV:BIST:TBAS ON is sent, this command overrides the Trigger state and sets it
to OFF. However, the 10 MHz remains out the Trig out port.
If the SERV:BIST:TBAS ON has been sent, the 10 MHz is on and the
OUTPut:TRIGger[:STATe] is then toggled to ON, the channel trigger is now
routed to the Trig out overriding the service command turning the 10 MHz to off.
If the command is set to:
– ON, the 10 MHz time base signal is sent to the front panel trigger out
– OFF, the 10 MHz time base signal is disabled.
Syntax
SERV
Example
SERV:BIST:TBAS:STAT OFF
624
connector.
:BIST
:TBAS
:STAT
This command disables the signal.
Keysight N8262A Programming Guide
Space
0|OFF
1|ON
?

Advertisement

Table of Contents
loading

This manual is also suitable for:

N8262a

Table of Contents