Keysight Technologies P Series Programming Manual page 369

Modular power meter
Hide thumbs Also See for P Series:
Table of Contents

Advertisement

[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2[:SELect] <string>
This command enters the name of the sensor calibration table or frequency
dependent offset table which is to be used. The CSET1 command selects the
sensor calibration table and the CSET2 command selects the frequency dependent
offset table. The calibration factor is interpolated from the table using the setting
for [SENSe[1]]|SENSe2:FREQuency.
If [SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe is set to OFF, the
NOTE
selected sensor calibration table or frequency offset table is not being used.
Syntax
SENS
1
SENS2
Parameters
Item
Description/Defaul t
string
String data representing a sensor calibration table, or
frequency dependent offset table name.
Keysight N8262A Programming Guide
:
:CSET
CORR
:CSET2
SENSe Subsystem
:SEL
1
Range of Values
Any existing table name
(Existing table names
can be listed using MEMory:CATalog:TABle?).
10
string
Space
?
369

Advertisement

Table of Contents
loading

This manual is also suitable for:

N8262a

Table of Contents