1
Power Meter Remote Operation
Using READ?
ABOR1
CONF1 DEF,3
READ1?
Some fine tuning of the above program segment can be carried out for example,
by setting the trigger delay off. The following program segment assumes that
Channel A is currently being measured on the upper window.
1 ABOR1
2 CONF1 DEF,3
3 TRIG1:DEL:AUTO OFF
4 READ1?
Using INITiate and FETCh?
The following program segment assumes that Channel A is currently being
measured on the upper window.
ABOR1
CONF1 DEF,3
INIT1
FETC1? DEF,3
72
Aborts Channel A
Configures the upper window to make a measurement
using the current setting of the expected power and
source list and a resolution setting of 3
Takes the upper window's measurement. This is Channel A
or B measurement depending on current window setup.
Aborts Channel A
Configures the upper window to make a measurement
using the current setting of the expected power and
source list and a resolution setting of 3
Causes Channel A to make a measurement
Retrieves the upper window's measurement
Keysight N8262A Programming Guide