–231
Data questionable;ZERO ERROR ChB
Power meter zeroing failed on Channel B. The most likely cause is attempting to zero when some power
signal is being applied to the power sensor.
–241
Hardware missing
The power meter is unable to execute the command because either no power sensor is connected or it
expects an E-Series power sensor and one is not connected.
–310
System error;Dty Cyc may impair accuracy with ECP sensor
This indicates that the sensor connected is for use with CW signals only.
–310
System error;Ch A Dty Cyc may impair accuracy with ECP sensor
This indicates that the sensor connected to Channel A is for use with CW signals only.
–310
System error;Ch B Dty Cyc may impair accuracy with ECP sensor
This indicates that the sensor connected to Channel B is for use with CW signals only.
–310
System error;Sensor EEPROM Read Failed - critical data not found or unreadable
This indicates a failure with your E-Series power sensor. Refer to your power sensor manual for details on
returning it for repair.
–310
System error;Sensor EEPROM Read Completed OK but optional data block(s) not found or unreadable
This indicates a failure with your E-Series power sensor. Refer to your power sensor manual for details on
returning it for repair.
–310
System error;Sensor EEPROM Read Failed - unknown EEPROM table format
This indicates a failure with your E-Series power sensor. Refer to your power sensor manual for details on
returning it for repair.
–310
System error;Sensor EEPROM < > data not found or unreadable
Where < > refers to the sensor data block covered, for example, Linearity, Temp - Comp (temperature
compensation).
This indicates a failure with your E-Series power sensor. Refer to your power sensor manual for details on
returning it for repair.
–321
Out of memory
The power meter required more memory than was available to run an internal operation.
–330
Self-test Failed;
The -330, "Self-test Failed" errors indicate that you have a problem with your power meter. Refer to
Contacting Keysight
Technologies on page 119 for details of what to do with your faulty power meter.
–330
Self-test Failed;Measurement Channel Fault
–330
Self-test Failed;Measurement Channel A Fault
Keysight N8262A Programming Guide