OUTPut:RECorder[1]|2:LIMit:LOWer <numeric_value>
This command sets the minimum scaling value for the specified recorder output.
The units used are dependent on the units currently set for the CALC block
specified in OUTPut:RECorder[1]|2:FEED <data_handle>.
Syntax
OUTP
:REC
Parameters
Item
numeric_value
Example
OUTP:REC:LIM:LOW –90
Reset Condition
On reset, the minimum scaling value is set to –150 dBm.
Keysight N8262A Programming Guide
1
2
Description/Defaul t
A numeric value for the minimum scaling value. The units
used—dBm, W or %—are dependent on the units currently
set for the CALC block specified in
OUTPut:RECorder[1]|2:FEED <data_handle>.
:LOW
:LIM
This command sets the minimum scaling
value to –90.
OUTPut Subsystem
Space
numeric_value
?
Range of Values
–150 to +230 dBm
1 aW to 100 XW
0 % to 999 %
8