Keysight PXA Signal Analyzers Option LNP Low Noise Path Option MPB Microwave Preselector Bypass Notice: This document contains references to Agilent. Please note that Agilent’s Test and Measurement business has become Keysight Technologies. For more information, go to www.keysight.com. Installation Note...
Page 2
The information contained in this document is subject to change without notice. Keysight Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
Page 3
Option LNP or Option MPB Installation Kit Option LNP or Option MPB Installation Kit Products Affected: PXA N9030A, Options 508, 513, 526, 543, 544, and 550 Serial Numbers: To Be Performed By: (X) Keysight Service Center (X) Personnel Qualified by Keysight...
Page 4
Option LNP or Option MPB Installation Kit 1. At the time of manufacture the hardware related to this option was fully adjusted and the option performance was verified to be within its warranted specifications. Within one year of the initial calibration date of the instrument this option is fully calibrated with no further adjustment or verification testing.
Option LNP or Option MPB Installation Kit Contents Keysight Part Quantity Description Number Installation Note This note Option Upgrade Entitlement Certificate 5964-5178 Tools Required — Personal computer with internet access and USB port — USB storage device with > 2 GB free memory If adjustments and performance testing is required: —...
Option LNP or Option MPB Installation Kit Figure 1 USB Storage Device Configuration Menu 7. The signal analyzer will automatically consume the License File. (This may take a few minutes) When the License File is consumed the Keysight License Manager will display a “Successful License Installation”...
Page 7
Option LNP or Option MPB Installation Kit Manually Control the Option Assure the signal analyzer is in Spectrum Analyzer mode. View the 4.8 GHz calibrator signal: Preset the analyzer by pressing the green Mode Preset key. Press Input/Output, RF Calibrator, 4.8 GHz. Press Freq/Channel, enter 4.8 GHz.
Page 8
For assistance, contact your nearest Keysight Technologies Sales and Service Office. To find your local Keysight office access the following URL, or if in the United States, call the following telephone number: http://www.keysight.com/find/assist...