Service:bist:calibrator <Boolean - Keysight Technologies P Series Programming Manual

Modular power meter
Hide thumbs Also See for P Series:
Table of Contents

Advertisement

SERVice:BIST:CALibrator <boolean>
This command enables/disables the calibrator self-test during power-up. It can be
used to disable the self-test if it incorrectly indicates a failure. If a load, for example,
a sensor, is connected to the calibrator port this could cause the self-test to fail.
Also, if it fails the self-test, a Pop-up is displayed for 5 seconds, stating -If Ref
Calibrator test fails disconnect any load attached to it and re-try test.
Syntax
SERV
Example
SERV:BIST:CAL OFF
Query
SERVice:BIST:CALibrator?
The query enters a 1 or 0 into the output buffer indicating the status of the
self-test.
– 1 is returned when the self-test is enabled
– 0 is returned when the self-test is disabled
Query Example
SERV:BIST:CAL?
Keysight N8262A Programming Guide
:BIST
:CAL
Space
0|OFF
1|ON
ONCE
?
This command disables the calibrator
self-test during power-up.
This command queries whether the self-test
is enabled or disabled.
SERVice Subsystem
16
615

Advertisement

Table of Contents
loading

This manual is also suitable for:

N8262a

Table of Contents