Power Meter Remote Operation
The following example uses the expected value parameter to specify a value of
–50 dBm. This selects the power meter's lower range (refer to
for details of the range breaks). The resolution parameter is defaulted, leaving it at
its current setting. The source list parameter specifies a Channel B measurement.
The measurement is carried out on the upper window.
Using READ?
ABOR2
CONF1 -50,DEF,(@2)
READ1?
Some fine tuning of measurements can be performed using the CONFigure and
READ? commands. For example, in the above program segment some fine tuning
can be performed by setting the filter length to 1024 and the trigger delay off.
1 ABOR2
2 CONF1 -50,DEF,(@2)
3 SENS2:AVER:COUN 1024
4 TRIG2:DEL:AUTO OFF
5 READ1?
Using INITiate and FETCh?
ABOR2
CONF1 -50,DEF,(@2)
INIT2
FETC1? -50,DEF,(@2)
70
Aborts Channel B
Configures the upper window to make a
Channel B measurement using an expected
power of –50 dBm and the current resolution
setting
Takes the upper window's measurement
Aborts Channel B
Configures the upper window to make a Channel
B measurement using an expected power of
–50 dBm and the current resolution setting
Causes Channel B to make a measurement
Retrieves the upper window's measurement
Keysight N8262A Programming Guide
"Range"
on page 86