CALCulate Subsystem

CALCulate[1]|2|3|4:LIMit:FCOunt?

This query returns the total number of limit failures for the specified window/
measurement.
If the appropriate STATe commands are set to ON, each time a measurement is
initiated on the specified window/measurement and the result is outside the
limits, the counter is incremented by one.
If the measured value is equal to a limit, this is a limit pass.
The counter is reset to zero by any of the following commands:
– *RST
– CALCulate[1]|2|3|4:LIMit:CLEar:IMMediate
– CALCulate[1]|2|3|4:LIMit:CLEar:AUTO ON
When CALCulate[1]|2|3|4:LIMit:CLEar:AUTO is set to ON, the counter is set to
zero each time a measurement is:
– measured using MEASure?
– read using READ?
– initiated using:
When CALCulate[1]|2|3|4:LIMit:CLEar:AUTO is set to ONCE, the counter is set
to zero the first time a measurement is:
– measured using MEASure?
– read using READ?
– initiated using:
The maximum number of errors is 2
the counter returns to zero.
202
– INITiate[:IMMediate] or,
– INITiate:CONTinuous ON
– INITiate[:IMMediate] or,
– INITiate:CONTinuous ON
16
16
–1. If more than 2
Keysight N8262A Programming Guide
–1 errors are detected