Initiate:continuous:sequence[1]|2 <Boolean - Keysight Technologies P Series Programming Manual

Modular power meter
Hide thumbs Also See for P Series:
Table of Contents

Advertisement

INITiate:CONTinuous:SEQuence[1]|2 <boolean>
This command sets the power meter for either a single trigger cycle or continuous
trigger cycles. A trigger cycle means that the power meter exits the wait for
trigger state and starts a measurement. When entering local mode,
INITiate:CONTinuous is set to ON.
If INITiate:CONTinuous:SEQuence[1|2] <boolean> is set to:
– OFF, the trigger system remains in the idle state until it is set to ON, or
– ON, the trigger system is initiated and exits the idle state. On completion of
This command performs the same functions as INITiate[1]|2:CONTinuous
NOTE
<boolean>.
Syntax
Example
INIT:CONT:SEQ2 ON
Keysight N8262A Programming Guide
INITiate:IMMediate is received. Once this trigger cycle is complete the
trigger system returns to the idle state.
each trigger cycle, the trigger system immediately commences another trigger
cycle without entering the idle state.
INIT
:CONT
1
:SEQ
2
This command places Channel B in a wait for
trigger state.
TRIGger Subsystem
Space
0|OFF
1|ON
?
14
575