10
SENSe Subsystem
[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:TIME <numeric_value>
This command sets the length of the time-gated period (time-gate length) for
time-gated measurements for the P-Series sensors which are set to NORMal
mode.
Syntax
Parameters
Item
numeric_value
Example
SENS2:SWE3:TIME 0.001
398
:
SENS
1
SENS2
Description/Defaul t
The length of the time gated period in seconds.
– DEF: the default value is 100 µs
Units are resolved to 1 ns.
: TIME
SWE
1
2
3
4
This command sets the length to 0.001
seconds.
Keysight N8262A Programming Guide
Space
numeric_value
DEF
?
Range of Values
0 to 1 second
DEF