Setting the Range
To set the range manually use the following command:
[SENSe[1]]|SENSe2:POWer:AC:RANGe <numeric_value>
If the <numeric_value> is set to:
– 0, the sensor's lower range is selected. (For example, this range is –70 to
– 1, the sensor's upper range is selected. (For example, this range is –14.5 to
For details on the range limits of other E-Series power sensor refer to the
appropriate power sensor manual.
For further information on this command refer to
To enable autoranging use the following command:
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO ON
Use autoranging when you are not sure of the power level you will be measuring.
Keysight N8262A Programming Guide
–13.5 dBm for the E4412A power sensor.)
+20 dBm for the E4412A power sensor.)
Power Meter Remote Operation