10
SENSe Subsystem
[SENSe[1]]|SENSe2:TRACe:OFFSet:TIME <numeric_value>
This command sets the delay between the delayed trigger point and the start of
the trace for P-Series sensors are set to NORMal mode.
Syntax
SENS
SENS2
Parameters
Item
numeric_value
Example
SENS:TRAC:OFFS:TIME TIME 0.05 This command sets the delay to 0.05
402
:
TRAC
1
Description/Defaul t
The length of the delay in seconds.
– DEF: the default value is 0 seconds
Units are resolved to 1 ns.
:OFFS
Space
:TIME
seconds.
Keysight N8262A Programming Guide
numeric_value
DEF
?
Range of Values
–1 to 1 second
DEF