TRIGger Subsystem
TRIGger[1]|2:DELay:AUTO <boolean>
This command is used to determine whether or not there is a settling-time delay
before a measurement is made.
When this command is set to:
– ON, the power meter inserts a settling-time delay before taking the requested
– OFF, the power meter makes the measurement immediately a trigger is
TRIGger[1]|2:DELay:AUTO is ignored if TRIGger[1]|2[:IMMediate] is used to
trigger the power meter.
Syntax
580
measurement. This settling time allows the internal digital filter to be updated
with new values to produce valid, accurate measurement results. The trigger
with delay command allows settling time for the internal amplifiers and filters.
It does not allow time for power sensor delay.
In cases of large power changes, the delay may not be sufficient for complete
settling. Accurate readings can be assured by taking two successive
measurements for comparison.
received.
TRIG
1
2
:AUTO
:DEL
Space
0|OFF
1|ON
?
Keysight N8262A Programming Guide