Figure 2-1 Measurement Display Configure Block Window; Optional Parameters - Keysight Technologies P Series Programming Manual

Modular power meter
Hide thumbs Also See for P Series:
Table of Contents

Advertisement

2
MEASurement Commands
Figure 2-1

Optional Parameters

CONFigure, FETCh?, READ? and MEASure? have the following three optional
parameters:
– An expected power value
– A resolution
– A source list
Expected Power Value
An <expected_value> parameter is only required if you are using an E-Series
power sensor. It has no effect on P-Series or 8480 Series Power Sensor. The value
entered determines which of the power sensor's two ranges is used for the
measurement. If the current setting of the power sensor's range is no longer valid
for the new measurement, specifying the expected power value decreases the
time taken to obtain a result.
126
CONfigure1?
upper window/upper measurement
CONFigure3?
upper window/lower measurement
CONFigure2?
lower window/upper measurement
CONFigure4?
lower window/lower measurement
Measurement display CONFigure block window
Keysight N8262A Programming Guide