Keysight Technologies P Series Programming Manual page 133

Modular power meter
Hide thumbs Also See for P Series:
Table of Contents

Advertisement

CONFigure[1]|2|3|4[:SCALar][:POWer:AC]
[<expected_value>[,<resolution>[,<source list>]]]
This command is used on the specified window/measurement to set:
– The expected power level of the measurement.
– The resolution of the window/measurement.
– The channel on which the measurement will be made.
Syntax
CONF
1
2
3
4
Parameters
Refer to
parameters in this command.
Item
expected_value
Keysight N8262A Programming Guide
:SCAL
:POW
expected_value
Space
DEF
"Optional Parameters"
Description/Defaul t
A numeric value for the expected power level. The units of
measurement are dBm and W. The default units are defined
by UNIT:POWer.
:AC
,
resolution
DEF
on page 126 for additional details on the
MEASurement Commands
,
source list
Range of Values
Sensor dependent.
[a]
DEF
2
133

Advertisement

Table of Contents
loading

This manual is also suitable for:

N8262a

Table of Contents