5.2 Status Screen

When the self test was successful, the status screens appear on the front panel displays.
You can adjust the screen brightness or screen contrast to your requirements from
Chromeleon or on the front panel displays (→ page 125 for the pump display and page 131
for the column compartment display).

5.2.1 Pump Module

The status screen for the pump module shows the following information:
Nano Pump
NC_Pump
0.35 ul/min
350 nl/min
%A: 85
This information is
displayed ...
Pump name
Flow on or
Flow off or
Purge
Flow rate
Pressure
%A, %B or
%A, %B, %C
Page 86
Flow on
Flow on
764 bar
764 bar
%B: 15
Fig. 42: Status screen (example) (here for the NCS-3500RS)
Description
The pump name is the name that you specified on the Pumps page, under the
related Pump Device Name, in the Properties dialog for the module in the
Chromeleon Server Configuration program (→ page 48).
Flow on appears when the pump delivers with the specified flow rate.
Flow off appears when the pump is not delivering.
Purge appears when a purge cycle is running.
When the pump is pumping, the display always shows the flow rate that the
pump is actually delivering. When a flow ramp has been set in Chromeleon, it
may take some time until the specified target flow is reached.
While is pump is idle, the nominal flow rate is displayed and the flow value is
flashing.
Shows the pressure in the unit that you specified on the Limits page in the
Properties dialog for the module in the Chromeleon Server Configuration
program (→ page 48).
Shows the components of the solvent in percent of the total flow.
Tip: If the purge cycle for the NC pump is started from Chromeleon, 'Purging'
appears on the display. In addition, the component is indicated for which the
purge cycle is running
LoadingPump
Flow on
20.0 ul/min
2 bar
%A: 100 %B: 0
%C: 0