Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual page 59

Nano/cap system. nano/cap pump
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
Columns Page
The Columns page shows all columns that are available for column monitoring (→ page 34).
Fig. 18: Columns page
Select the columns for which you want to use column identification by the check box in the
Enabled column. If a check box is cleared, the column will not be available in
Chromeleon. The Name column displays the name used to identify the column in the
installation environment and in the Chromeleon Client program. To use existing control
panels, accept the default name.
To change a column name, overwrite the existing name directly in the corresponding line. If
you enter a different name, you may have to re-link the controls on the control panels and edit
the column name in the program files.
The information about the column is continuously updated and can be reviewed at any time in
Chromeleon (→ page 129).
Signals Page
The Signals page lists all available signals. Select a check box to enable the signal. If a
check box is cleared, the signal will not be available in Chromeleon. To change a signal
name or the scaling factor, overwrite the existing entry directly in the corresponding line.
The NC pump pressure signal is selected by default. Accept this setting if you want to record
the column pressure of the NC pump. If a signal channel is selected, Chromeleon generates
the channel for data acquisition.
Depending on the flow meter that is installed, make sure that the corresponding signals are
selected in order to gain information from the signal channels on the respective flow meter
performance.
For more information, see section 5.5.6 (→ page 121).