Pin
Signal Name
1
2
Solvent Rack Error
3
4
Solvent Rack Leak
5
6
V_Degas
7
GND_Degas
8
VCC
9
10
GND
11
GND
12
GND
13
14
V_Degas
15
GND_Degas
Page 256
Signal Level
+15V_supply
Ground_supply
+15V_supply
Ground_supply
Fig. 97: 15-pol. Solvent Rack port (female)
Remark
Reserved
TTL_high with solvent rack errors
Jumper to pin 9
TTL_high with Solvent Rack leaks
Reserved
Supply for the solvent rack
Reference potential for V_Degas
Voltage for logic devices +5V
Jumper to pin 3
Reference potential for VCC
Reference potential for VCC
Reference potential for VCC
Reserved
Supply for the solvent rack
Reference potential for V_Degas