Combinations Of Ultimate 3000 Pumps And Solvent Racks - Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual

Nano/cap system. nano/cap pump
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS

2.2.2 Combinations of UltiMate 3000 Pumps and Solvent Racks

The Solvent Racks of the UltiMate 3000 system series are an ideal complement to the
module, whether you need high efficiency degassing of the solvents or simply want to
safely organize your solvent reservoirs.
Solvent Rack
SRD-3200 Solvent Rack with analytical 2-channel vacuum degasser
SRD-3400 Solvent Rack with analytical 4-channel vacuum degasser
SRD-3600 Solvent Rack with analytical 6-channel vacuum degasser
SR-3000 Solvent Rack without degasser
In an UltiMate 3000 RSLCnano system, a SRD-3400 Solvent Rack with analytical 4-
channel vacuum degasser should be used together with the NCS-3500RS.
Tip: If the UltiMate 3000 system includes an UltiMate 3000 series autosampler, you
should degas also the wash solution on a continuous basis, by using the
vacuum degasser in the Solvent Rack. The procedure how to prepare and
install the liquid lines for the wash solution is similar to the steps for the
solvent supply lines (→ page 63). For information about online degassing of
the wash solution, see the Autosampler manual.
Page 20
Part No.
5035.9250
5035.9245
5035.9230
5035.9200
Operating Instructions

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ncp-3200rsNcs-3500rs

Table of Contents