Thermo Scientific TriPlus 300 Preinstallation Manual
Thermo Scientific TriPlus 300 Preinstallation Manual

Thermo Scientific TriPlus 300 Preinstallation Manual

Headspace sampling system

Advertisement

Quick Links

Thermo Scientific
TriPlus 300 Headspace
Sampling System
Preinstallation Requirements Guide
P/N 317 09672 Revision C December 2015

Advertisement

Table of Contents
loading

Summary of Contents for Thermo Scientific TriPlus 300

  • Page 1 Thermo Scientific TriPlus 300 Headspace Sampling System Preinstallation Requirements Guide P/N 317 09672 Revision C December 2015...
  • Page 2 © 2015 Thermo Fisher Scientific Inc. All rights reserved. TriPlus 300 Headspace, TRACE 1300, TRACE 1310, ISQ Series, TSQ 8000 Series, ITQ, and DSQ II are trademarks of Thermo Fisher Scientific Inc., and its subsidiaries. Published by Thermo Fisher Scientific S.p.A., Strada Rivoltana 20090 Rodano-Milan, Italy Tel: +39 02 95059303;...
  • Page 3 300 Headspace Installation Request Form Once you read the TriPlus 300 Headspace Preinstallation Requirements Guide, print and complete this form. After all the requirements on this form are fulfilled, sign and date the form. Then mail or fax this form to your local Thermo Fisher Scientific sales/service office.
  • Page 5 Offices for Thermo Scientific CMD Products For up-to-date contact information, visit www.thermoscientific.com. North America France (Also representing French-speaking North Africa, United States Algeria, Morocco, and Tunisia) 1400 North Point Pkwy #10 16 Avenue du Québec West Palm Beach, FL 33407...
  • Page 6 Offices for Thermo Scientific CMD Products—Continued Europe—continued Building 6 No.27 Xin Jinqiao Road Pudong District Sweden, Norway, and Finland Shanghai 201206 Pyramidbacken 3 Phone......[86] (21) 6865 4588 S-141 75 Kungens Kurva (Stockholm) Fax ......[86] (21) 6445 7830 Sweden E-mail: service.sid.nordic@thermofisher.com Phone......[46] (0) 8 556 468 20...
  • Page 7 Declaration Manufacturer: Thermo Fisher Scientific Thermo Fisher Scientific is the manufacturer of the instrument described in this manual and, as such, is responsible for the instrument safety, reliability and performance only if: • installation • re-calibration • changes and repairs have been carried out by authorized personnel and if: •...
  • Page 8 Notice on Lifting and Handling of Thermo Scientific Instruments For your safety, and in compliance with international regulations, the physical handling of this Thermo Fisher Scientific instrument requires a team effort to lift and/or move the instrument.
  • Page 9 WEEE Compliance This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2012/19/EU. It is marked with the following symbol: Thermo Fisher Scientific has contracted with one or more recycling or disposal companies in each European Union (EU) Member State, and these companies should dispose of or recycle this product.
  • Page 10 Conformità RAEE Questo prodotto è marcato con il seguente simbolo in conformità alla direttiva europea 2012/19/EU (RAEE) sui rifiuti di apparecchiature elettriche ed elettroniche: Thermo Fisher Scientific si è accordata con una o più società di riciclaggio in ciascun Stato Membro della Unione Europea (EU), e queste società...
  • Page 11: Table Of Contents

    Chapter 1 TriPlus 300 Headspace Site Preparation ........1 Entrance Requirements .
  • Page 12 Contents TriPlus 300 Headspace Preinstallation Requirements Guide Thermo Scientific...
  • Page 13: Preface

    Preface This guide contains preinstallation requirements that must be in compliance before installing the TriPlus 300 Headspace sampling system. About Your System Thermo Fisher Scientific systems operate safely and reliably under carefully controlled environmental conditions. If the equipment is used in a manner not specified by the manufacturer, the protections provided by the equipment might be impaired.
  • Page 14: Contacting Us

    Thermo Fisher Scientific office, or affiliate GC-GC/MS Customer Support. Related Documentation In addition to this guide, Thermo Scientific provides the following documents for the TriPlus 300 Headspace. • TriPlus 300 Headspace Preinstallation Requirements Guide, PN 31709672 • TriPlus 300 Headspace User Guide, PN 31709671...
  • Page 15: Safety Symbols And Signal Words

    HARMFUL. Indicates that the presence of harmful material will, could, or might occur. INSTRUMENT DAMAGE: Indicates that damage to the instrument or component might occur. This damage might not be covered under the standard warranty. Thermo Scientific TriPlus 300 Headspace Preinstallation Requirements Guide...
  • Page 16: Instrument Markings And Symbols

    Instrument Markings and Symbols Table 1 explains the symbols used on Thermo Fisher Scientific instruments. Only a few of them are used on the TriPlus 300 Headspace sampling system. See the asterisk. Table 1. Instrument Marking and Symbols (Sheet 1 of 2) Symbol...
  • Page 17: Safety Information And Warnings

    Thermo Fisher Scientific representatives during installation, and repair of the TriPlus 300 Headspace sampling system, or parts of it (following the life-cycle principle), as well as for the end user TriPlus 300 Headspace sampling system in the lab during the learning phase and in routine work.
  • Page 18: General Considerations

    Preface Safety Information and Warnings General Considerations • Before using your system, consult the TriPlus 300 Headspace User Guide and related documents under all circumstances. • Changes or modifications to this unit not expressly approved by the party responsible for compliance, could void your’s authority to operate the equipment.
  • Page 19 To avoid damaging electrical parts, do not disconnect an electrical assembly while power is supplied to the TriPlus 300 Headspace sampling system. After the power is turned off, wait approximately 30 seconds before you disconnect an assembly.
  • Page 20: Other Hazards

    There is danger of crushing to fingers and hands. To avoid injury keep your hands away from moving parts during operation. Turn off the power to the TriPlus 300 Headspace sampling system if you must reach inside a mechanically powered system with moving parts.
  • Page 21: Working With Toxic Or Other Harmful Compounds

    Material Safety Data Sheet (MSDS). Use personal protective equipment according to the safety requirements. The TriPlus 300 Headspace sampling system requires the use of several chemical products with different hazard characteristics, which are present in vials and syringes. Before using these substances or replacing the syringe, please read the hazard indications and information reported in the MSDS supplied by the manufacturer referring to the relevant CAS number.
  • Page 22: Maintenance

    (color) of the instrument, table or other nearby objects. If in doubt, please contact your Thermo Fisher Scientific representative to verify the compatibility of the type or composition of solvents with the TriPlus 300 Headspace sampling system.
  • Page 23: Triplus 300 Headspace Site Preparation

    You may find all the instruction in the Preinstallation Requirements Guide of your GC-GC/MS system. The TriPlus 300 Headspace sampling system operates reliably under controlled environment conditions. Operating or maintaining a system outside the specifications outlined in this guide might cause many different types of system failures.
  • Page 24: Entrance Requirements

    (*) Including pallet Space and Load Requirements TriPlus 300 Headspace sampling system must be installed on the left of the GC-GC/MS units. Use the following tables to verify and determine the space and mass requirements for the instrument of your GC-GC/MS system.
  • Page 25: Gc-Gc/Ms System + Triplus 300 Headspace Configuration Space And Mass Requirements

    GC-GC/MS System + TriPlus 300 Headspace Configuration Space and Mass Requirements TriPlus 300 Headspace sampling system must o be installed on the left of the GC-GC/MS system. The maximum distance from the GC-GC/MS system must considers the length of the transfer line that connects the TriPlus 300 Headspace sampling system to the GC injector.
  • Page 26: Lighting Requirements

    • 5 cm between a GC and a DSQ II mass spectrometer A distance of about 3/5 cm between the GC-GC/MS system and the TriPlus 300 is left for opening the top cover of the Main through the two lateral push-buttons.
  • Page 27: Power Line

    Poor grounding represents a danger for the operator and might seriously affect the instrument performance. Do not connect the TriPlus 300 Headspace sampling system to lines feeding devices of a heavy-duty nature, such as motors, UV lamps, refrigerators, air compressors, and other devices that can generate disturbances.
  • Page 28: Environment Requirements

    The operating environment in your laboratory is affected by such factors as temperature, humidity, particulate matter, and electrostatic discharge. It is your responsibility to provide an acceptable operating environment for your TriPlus 300 Headspace sampling system. Attention to the operating environment will ensure continued high performance of your TriPlus 300 Headspace sampling system.
  • Page 29: Gas Equipment Requirements

    You will need a supply of ultra-high purity GC gases. The TriPlus 300 Headspace sampling system requires auxiliary gases for all the operating processes, and receives the carrier gas from the GC unit used to transfer the sample to the injector.
  • Page 30 Connectors will vary with cylinder size. Confirm that your regulator will work with your gas tank. All connections to the TriPlus 300 are 1/8 in. Swagelok fittings. TriPlus 300 Headspace Preinstallation Requirements Guide...
  • Page 31: Lan Network Requirements

    LAN Network Requirements The connection between the TriPlus 300 Headspace instrument and a Thermo Scientific Chromatography Data System (Chromeleon, Xcalibur, Chrom-Card, ChromQuest) must be carried out via Local Area Network (LAN).
  • Page 32: Receiving Instruments

    (static) IP address assigned to it. Receiving Instruments When you receive the TriPlus 300 Headspace instrument: 1. Inspect the boxes for damage on arrival. Our instruments are shipped by electronic equipment carriers who specialize in the handling of delicate equipment. Occasionally, however, equipment is inadvertently damaged in transit.

Table of Contents