7.6.3.3 Installing the Piston Seals and Reassembling the Pump Head
Description
Main piston seal (reversed phase)
(→ Fig. 65, page 184, no. 7 )
NC pump and loading pump (2 seals)
Support ring (→ Fig. 65, no. 6) for
NC pump and loading pump
Tip: Never reassemble the pump head with bare hands. Put on a pair of clean room
gloves to prevent contamination on the pump parts. Even minute particles may
cause damage to the system and result in poor pump performance.
1. The support ring is not a wear part and thus, you need not replace the support ring
when you replace the piston seal. However, before reinstalling the support ring, you
should clean the support ring from the inside and outside, by using isopropanol, a
cleaning swab (part no. 6040.0006), and a lint-free cloth or tissue.
2. Pipette a few drops of isopropanol onto the edge in the pump head on which the piston
seal will rest.
3. First, slide the support ring, and then slide the piston seal onto the seal insertion tool
(→ Fig. 70, page 195). Mind the correct orientation of the seal. The open side of the
seal must face away from the tool.
This side faces away
from the tool.
4. Push the piston seal tool into the pump head as far as it goes in.
Operating Instructions
Fig. 72: Correct orientation of the piston seal
UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
Part No.
6266.0305
6040.0012
Page 197