7 Service

7.1 General Notes and Safety Precautions

The following sections describe all service and repair procedures that the user may
perform. All other maintenance and service procedures must be performed only by Thermo
Fisher Scientific service personnel
Warning:
Avertissement: Les composants fluidiques de l'instrument peuvent être remplis de
Before starting maintenance or service procedures, observe the following precautions:
For all service and repair procedures, observe all precautionary statements provided in
these operating instructions.
NC pump
When the fluid components of the NC pump are filled with liquid and the solvent
reservoirs are located above the pump outlet during pump operation, the hydrostatic
pressure in the system may cause eluent to escape when you open a fluid connection in
the NC pump. Before you open a fluid connection, position the reservoirs below the
connection to be opened. Empty the solvents lines. If solvent shut off valves are installed
on the solvent lines, you can close the shut off valves instead.
Column Compartment
Do not touch any metal or plastic parts inside the column chamber while the temperature
is higher than 50 °C. Wait for the chamber to cool down before you carry out any work in
the column chamber. To shorten the cool-down time, consider setting a lower
temperature and opening the front panel door.
Operating Instructions
The fluid components of the device may be filled with solvents that
are harmful to health. Wear appropriate personal protective
equipment. Rinse the fluid components with an appropriate solvent
to remove harmful substances.
For information about the proper handling of a particular substance
and for advice on specific hazards, refer to the material safety data
sheet for the substance you are using. Observe the guidelines of
Good Laboratory Practice (GLP).
solvants nocifs. Portez l'équipement de protection personnel
approprié. Rincez les composants fluidiques avec un solvant
approprié afin d'éliminer les substances nocives.
Pour les informations sur la manipulation correcte des composés et
des recommandations pour les situations de risque spécifiques,
veuillez consulter la fiche de données de sécurité des substances que
vous utilisez. Veuillez respecter des directives des Bonnes Pratiques
de Laboratoire (BPL).
UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
Page 175