Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual page 57

Nano/cap system. nano/cap pump
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

Flow
Shows the allowed flow range.
Enter a new value in the Minimum and/or Maximum box to change the limits for the
flow rate within the allowed range.
On the Limits (Loading) page, under Flow Unit, you can select the unit in which the
loading pump flow is displayed (µl/min or ml/min).
Pressure
Shows the allowed pressure range.
Enter a new value in the Minimum and/or Maximum box to change the pressure limits
within the allowed range.
Click the arrow of the Pressure Unit box and select a different pressure unit (psi, bar, or
MPa) on the list if required.
These values are the absolute limits. Within these limits, you can set specific limits on a
control panel or in a program file.
Solvents Page
The loading pump and the NC pump of the NCS-3500RS have separate Solvents pages.
Check and change the number of solvents delivered by the pump and the solvent names if
necessary.
Number of Solvents
Indicate how many solvents are delivered by the pump. The pump type determines how
many solvents can be delivered.
Solvent Names
Enter a name for each connected solvent. A solvent name can contain a maximum of
30 characters. The names of the solvents appear, for example, in the gradient display of
the online control panel and in the report.
Tip: Changing the solvent names on the Solvents page does not set up the flow
meter to operate properly with the respective solvent. Make sure that the
flow meter is also configured for the solvents that you use
(→ section 4.5.2, page 69).
Operating Instructions
Pump Type
NC pump
Loading pump
NCS-3500RS and NCP-3200RS
Max. Number of Solvents
2
3
UltiMate 3000 Series:
Page 49