6.7 Checking the Compression Values (Loading Pump)

Relevant only for the loading pump
Tip: The following applies only when the loading pump has backpressure.
The compression value of the pump head can provide valuable information for
troubleshooting. If you observe pump pulsation and/or shifts in retention times, check the
compression values:
On the pump display, select the Diagnostics menu and Compression.
In Chromeleon, open the Commands dialog box for the loading pump and select
LoadingPump_Wellness_RightBlock and Compression.
The values are indicated in percent; they show the compression of the last stroke. The
compression value should be lower than 100% and remain stable. If the value is close to
100%, the required precompression may not be reached and pulsation may occur.
The table shows guides values for some solvents:
Solvent
Acetonitrile
Methanol
Water
* with 100 % of degassed solvent delivered at 40 MPa, pump purged and stable for some time The values
are linear to the pressure.
If the compression value is lower than indicated in the table, the valve cartridge (outlet
check valve) may be defective. Replace the cartridge (→ page 182).
If the compression value is higher than indicated in the table
♦ The piston seals may be defective. Replace the piston seals (→ page 194).
♦ The valve cartridge (inlet check valve) may be defective. Replace the cartridge
(→ page 182).
If the compression is very high and the pressure is low, there may be air bubbles in the
system. Purge the loading pump (→ page 76).
Page 174
Compression*
70 % ± 10 %
75 % ± 10 %
50 % ± 10 %