3. Enter the name of the solvent and the viscosity. Note the tips further down in this
section.
♦ To add a new entry
Click into the line marked by the asterisk (*) in the leftmost column and type the
name and viscosity value in this line.
♦ To edit an entry
Click the entry that you want to edit and edit the entry as needed. Afterward, click
another entry.
Observe the following:
♦ The solvent name can contain any of the following characters a-z, A-Z, 0-9,
underscore ('_'), and percent (%).
♦ The name must not start with a digit.
♦ The name must be unique.
♦ Avoid entries with different names but identical viscosities values. Combine these
entries, instead (→ Fig. 55).
Chromeleon always uses the entry for a specific viscosity that appears first on the
viscosity list (→ Fig. 56). Therefore, if more than one entry exists for a specific
viscosity, the Chromeleon report may not show the name of the solvent that is
actually used but the name of the solvent with the specified viscosity value found
first on the viscosity list.
Example:
The viscosity list includes among others the following entries: Solvent2__120_0%
and Solvent3__120_0%. The report will show Solvent2_120_0% as solvent name
even if you have selected Solvent3__120_0% on the viscosity list.