2. Start the Server Configuration program (→ page 45).
3. Right-click the module in the timebase and click Properties on the menu.
4. On the General page (→ page 46), the firmware version provided by Chromeleon for
the module is displayed in the Firmware box. If more than one firmware version is
available for the module in Chromeleon, select the version you wish to transfer from
the Firmware list.
5. Click Download. A message displays the firmware version that is currently installed
in the module and the version that will be downloaded from Chromeleon.
Tip: If the module comes with a newer firmware than the version included in
Chromeleon, do not downgrade the firmware. Older firmware may be
incompatible with new hardware revisions.
6. Click Yes to start the download. (Click No to cancel the action.)
The download may take several minutes. The download is complete when Firmware
download completed successfully appears in the Messages Server window in the
Chromeleon Server Configuration program. The message appears also in the Chromeleon
Audit Trail.
Immediately after the new firmware has been downloaded from Chromeleon to the module,
the module performs a reset. For about 15 seconds, the internal boot loader is updated. Do not
turn off the module while the boot loader is updating.
If the download from Chromeleon is not successful, the related messages appear in the Audit
Trail. In this case, turn the module off and on again. Repeat the download as described above.
If the download fails again, contact Thermo Fisher Scientific Service for Dionex HPLC
Products.