3.4.3 Connecting the Power Cord

Use the power cord shipped with the module to connect the module to the main power
source. Connect the power cord from the main power receptacle on the rear panel
(→ Fig. 2, page 24) to a grounded power source. No manual adjustment is required to
adapt the line voltage to local voltage requirements.
Warning:
Avertissement: Utilisez uniquement les cordons d'alimentation électrique spécifique

3.4.4 Connecting a Solvent Rack

If the UltiMate 3000 system includes a SRD-3x00 Solvent Rack, connect the Solvent
Rack port on the rear panel of the module to the 15-pin D-Sub connector on the rear panel
of the Solvent Rack. The appropriate connection cable is included in the accessories kit of
the Solvent Rack.

3.4.5 Connecting Devices (Mass Spectrometers) to the Digital I/O-Port

To connect a device, for example, a mass spectrometer, to a digital I/O port on the rear
panel of the module, use a 6-pin mini-DIN signal cable (part no. 6000.1004). The cable is
available from Thermo Fisher Scientific.
1. Plug the 6-pin connector of the mini-DIN cable into the Digital I/O port 1 (or 2). For
information about the functions of the connector pins and pin assignment, see
page 255.
2. For each relay output or digital input to be used, connect the appropriate signal wire
and ground wire to the corresponding connectors on the device (→ documentation
provided with the device).
For example, if you want to use a relay output to control a mass spectrometer, use the
wires 5 (brown) and 6 (white).
3. When setting up the module in the Chromeleon Server Configuration program
(→ page 46), select the corresponding relay outputs on the Relays page and the digital
inputs on the Inputs page.
The relay outputs and digital inputs are then available in Chromeleon, for example, in the
Commands dialog box under PumpModule, and can be programmed as required.
Page 42
Never use a power cord other than the power cords provided for the
device.
Do not use multiple sockets or extension cords. Using defective
multiple sockets or extension cords may cause personal injury or
damage to the device.
à l'instrument.
N'utilisez pas des blocs multiprise ou des câbles prolongateurs. Cela
pourrait entraîner des blessures corporelles ou endommager
l'instrument.
Operating Instructions